Data output – HEIDENHAIN ND 286 User Manual
Page 4
Data Output
There are four ways to output data:
➤ Press the MOD key until the PRINT indicator starts blinking (only possible with
"slow" data output), and start data output with the ENT key; or
➤ Send measuring data to the data output periodically; or
➤ Input a latch command over the D-sub connection EXT; or
➤ Input a latch command over the BCD connection.
Interface mode (see operating parameter P53)
Slow:
Output display values
Fast:
Output instantaneous values referenced to datum 1
(MIN/MAX/DIFF display values are not output)
A connecting cable (to a PC, for example) is available from HEIDENHAIN
(Id.-Nr. 206 420 ..); cable length up to 10 m (32.8 ft).
Operating parameters for data output: P23, P53 to P57
“ AMP-CHAMP“ connection (36-pin, female)
Pins
Assignment
Pins
Assignment
2
0
2
1
2
2
2
3
1
2
3
4
Decade 1
33
Sign
5
6
7
8
Decade 2
34
Ready
9
10
11
12
Decade 3
35
Meas. val. output
13
14
15
16
Decade 4
36
0V
17
18
19
20
Decade 5
21
22
23
24
Decade 6
25
26
27
28
Decade 7
29
30
31
32
Decade 8
Output levels
Low: U
≤
0.4 V with I
≤
6 mA High: U
≥
3.8 V with I
≤
2.6 mA
The output signals are TTL-compatible.
Latch levels
Low: U
≤
0.9 V with Imax
≤
6 mA
High: U
≥
3.9 V; or
TTL levels (internal 10 k
Ω
pull-up resistor).
Signal transit times
The following table lists approximate signal transit times. If you use the slow
data output and run functions such as measuring series or inch display at the
same time, the actual transit times can be twice as long as those listed here.
Concurrent data output (P55 )
Mode
P53
Latch time
Data output after
Fast
P54
Value from P54 / 2
Slow
t
≤
30 ms
t
≤
8 ms
Data output after external latch (P55 or )
Mode
P53
Min. pulse duration
Measured value stored after
Pulse / BCD
Contact
BCD
Pulse
Contact
Fast
3 µs
7 ms
0.3 µs
1.1 µs
4.8 ms
Slow
t
≥
8 ms
t
≥
13 ms 0.3 µs
1.1 µs
4.8 ms
Data output
Latch again after
Pulse/BCD Contact
Fast
≤
0.3 µs after internal latching
3 µs
7 ms
Slow
≤
7.5 ms after internal latching
3 µs
7 ms