Switch-on, Setting the datum – HEIDENHAIN ND 286 User Manual

Page 2

background image

Select datum 1 or 2.

Enter a value, such as 40.

You can switch from one datum to the other at any time.
Use datum 2 when you want to display incremental values.

4 0

Confirm entered value.

The ND 286 position display unit is designed for use with HEIDENHAIN linear or
rotary encoders with square-wave output signals. These encoders have one or more
reference marks, which may also be

distance-coded.

When a reference mark is crossed over, it generates a signal identifying that position
as a reference point. After switch-on, simply crossing over the reference mark
restores the relationship between axis positions and display values as it was last
defined by datum setting.

With distance-coded reference marks, a maximum traverse of only 20 mm suffices to
re-establish the datum.

Switch-On

0 ➤1

Ent...CL

Switch on reference mark evaluation.
• Display shows the value last assigned to the

reference mark position.

• REF indicator glows.
• Decimal point blinks.

Cross over reference mark.
Move the axis until the display becomes active and the
decimal point no longer blinks.

5 , 6 9 7

Turn on the power (switch located on rear panel).
• Display shows

.

• REF indicator blinks.
• Data interface shows ERROR 07

If you do not wish reference mark evaluation, press CL instead of ENT.

Setting the Datum

The datum setting procedure assigns a display value to a specific axis position.
The ND 286 allows you to set two independent datums.