Watson-Marlow 505Di User Manual
Page 5

5
•
Change direction by pressing the CW/CCW key. Indication of direction of rotation is provided via the LCD display.
•
Select the maximum speed : press the
Ù key and the Max key together. Select the minimum speed: press the Ú
key and the Max key together.
•
Press Start to start the pump. Press Stop to stop the pump.
Network
This facility gives full pump functionality under RS232 closed loop control via the 9 pin D connector. Up to 16 pumps can
be linked whilst still retaining individual pump control by using lead PR 0024. A network kit is available from Watson-
Marlow which includes Pumpnet 2, a DOS compatible control program and leads.
Step to Network in the Main menu and press Enter. The pump will now be under RS232 control. The keypad Stop key
will act as an emergency stop and disable RS232 settings if pressed.
5
1
9
6
Connections for RS232 signals (viewed from inside the plug)
RS232 cabling shown for CTR handshake
Use only twin shielded RS232 cables.
RS232 settings
Baud = 9600 ; Stop bits = 2 ; Data bits = 8 ; Parity = None ; Handshake = CTR or None; Auto echo = On
Pin
Function
Pin
Function
1
-
6
-
2
RX (Receive data)
7
RTS (Request to send)
3
TX (Transmit data)
8
CTS (Clear to send)
4
-
9
-
5
GND (Ground)
The following codes will operate the 505Di under RS232 control. They must be directed to the pump from a computer
serial port (or equivalent). Always terminate each command with a RETURN (ASCII CHR13).
nSPxxx
Load speed setting xxx to pump number n
nSI
Increment speed by 1rpm for pump n
nSD
Decrement speed by 1rpm for pump n
nGO
Start pump number n
nST
Stop pump number n
nRC
Change rotation direction for pump n
nRR
Set clockwise direction for pump n
nRL
Set anti-clockwise direction for pump n
nDOxxxxx,yyy
Set dose for pump number n in tachometer pulses (note 3)
nRS
Show status for pump number n (note 4)