Watson-Marlow 505Di User Manual
Page 33

33
English
I3. Indication of volume/weight dosed and number of doses remaining
J3. Indication of volume/weight dosed and number of doses/hour
K3. Option to zero the cumulative volume register
L3. Decision to stop dosing
M3. On-the-fly (mid batch) re-calibration facility - Pump will continue to dispense
N3. Enter actual volume dosed and pump will re-calibrate itself over 5 doses
O3. Indication of dose volume/weight and number of doses in program called
P3. Verification of time interval between doses and pump speed
Q3. Accept called program for dose or choose another program from memory
R3. Decision to accept program
S3. Move to next program in memory
T3. Decision to delete program from memory
U3. Option to delete program from memory
V3. Decision to not delete program
W3. Decision to delete program
X3.
Program is deleted
English
Error Messages
Progs and setup lost
Press enter
This screen indicates corruption of the RAM or that the
EPROM has been changed. In either case pressing Enter
will re-initialise the RAM and clear the error.
Check drive and
setup now
This screen indicates that the motor has stalled or that a
loss of tachometer pulses has been selected. Clear the
cause of stalling and power the pump off and on. If the error
persists seek qualified assistance.