Troubleshooting – Tecfluid MX4 For FLOMAT Series User Manual

Page 54

background image

54

15

TROUBLESHOOTING

TROUBLESHOOTING

Problem

Probable cause

Solution

The screen shows
empty pipe

Pipe is empty

Make sure that the pipe is
completely full, for example,
installing the flowmeter in a
vertical pipe with upwards flow

Isolation of the electrodes

Clean the sensor electrodes.

Electrode cable disconnected

Connect the cable between the
sensor and the electronic
converter

Liquid with very low conductivity

The flowmeter is not adequate for
the application

The flow rate is
unstable

Dirt on the electrodes

Clean the sensor electrodes

The product contains air or non-
conductive particles in suspension

Verify that the flowmeter is
adequate for this application

The flow rate displayed
is 0

Coil cable disconnected

Connect the cable between the
sensor and the electronic
converter

The flow rate is smaller than
programmed as CUT OFF

Decrease the value of the cut off
(see page 36)

The instrument displays
a value when there is
not flow

The sensor is damaged due to
electrodes corrosion

Electrode material not adequate
for the liquid

The displayed flow rate
is higher than expected

The electrodes are immersed but
the pipe is not completely full

Make sure that the pipe is
completely full, for example,
installing the flowmeter in a
vertical pipe with upwards flow

Display is blank

Blown fuse

Change the fuse

The analog output gives
always 4 mA or 20 mA

Current output range not properly
programmed

Program the range properly (see
page 40)

The analog output gives
always 0 mA

Cable disconnected

Check the cable connection

15