3 converter factor, 4 nominal diameter, 5 diagnosis – Tecfluid MX4 For FLOMAT Series User Manual
Page 30: 1 parameters
![background image](https://www.manualsdir.com/files/807348/content/doc030.png)
30
6.5.1 Parameters
This screen allows the diagnosis of the coil current, the differential voltage on the sensor
electrodes, and the conductivity of the liquid.
6.5
Diagnosis
When the flowmeter is not working properly, it can be useful to test some parts of the
instrument in order to locate the problem.
In this section a list with two possible options appears.
6.3 Converter
factor
The electronic converter factor is shown. It should coincide with the Fe parameter on the
converter label.
In the first row, a possible damage of the sensor coils can be detected. If there is a
symbol
Ѵ on this row, it means that the coils do not have any damage. If the symbol is X
it means that the coils are damaged. In this case, please contact your distributor.
6.4 Nominal
diameter
The value of the nominal diameter is always the internal diameter of the pipe.
6
INSTALLATION PARAMETERS