Tecfluid XT5 User Manual
Page 19

9 TECHNICAL
CHARACTERISTICS
9.1 Power
supply
230, 240, 115, 24 VAC 50, 60 Hz . 24 VDC.
Power consumption: ≤ 5 VA
9.2 Analog
output
4-20 mA. Active or passive. Galvanically isolated from the power supply.
9.3 Pulse
output
Opto-isolated. Vmax: 30 VDC. Imax: 30 mA.
Maximum frequency in mode “P/U” : 6.25 Hz
Maximum frequency in mode “Hz” : 10 000 Hz
Minimum frequency in mode “Hz” : 0.04 Hz
Optional: Vmax: 240VAC / 350 VDC. Imax: 100 mA
Maximum frequency in mode “P/U” : 6.25 Hz
Maximum frequency in mode “Hz” : 75 Hz
9.4 Totalizer
Nº of digits:
7 (2 decimals)**
Digit size:
8 mm
Reset:
By means of key
9.5
Flow rate indication
Nº of digits:
4 (up to 2 decimals configurable)**
Digit size:
5 mm
** When the available digits are full and the integers overflow a decimal is automatically
lost.
9.6 General
characteristics
Protection rating:
IP67
Ambient temperature range:
0 ... +60 ºC
9.7
Electrical characteristics referred to the analog loop and communications
Reception
impedance:
Rx
>
8,5
MΩ
Cx < 200
pF
This equipment complies with the following EEC directives:
Low voltage (73/23/CEE)
Electromagnetic compatibility (89/336/CEE)
In this case, applying the formula, the maximum frequency at the output is:
As the value exceeds 6.25 Hz, this factor can not be applied.
In this case there would be 2 solutions:
1. To change the factor to 1 pulse per litre, becoming the maximum frequency 100 times
smaller, fmax = 2.639 Hz.
2. To change the units of volume to m3. With this, the maximum flow rate is divided by
1000 and therefore the maximum frequency becomes fmax = 0.2639 Hz.
19
Hz
263.9
100
2.639
f
max