Tecfluid XT5 User Manual
Page 12

In the first screen the flow rate for the 4 mA point (lower range) is configured. In the next
screen the flow rate for the 20 mA point (upper range) is configured.
To select the number of decimals it must be taken into account that the instrument has 4
digits for flow rate indication. If two decimals have been selected, these will be seen
whilst the flow rate is not greater than 99.99. Above this value the indication will
automatically change to one decimal, and when the flow rate is greater than 999.9 the
indication will be done without decimals.
If one decimal is selected, the flow rate indication will have a maximum of one decimal.
If indication without decimals is selected, the flow rate will always be shown without
decimals.
For the selection of the flow rate units and the number of decimals it must be taken into
account that an indication with an excess of decimals may give the sensation of instability
of the reading. As a general rule it can be considered that the reading should not have
more than a total of 3 digits (integer + decimals).
3.2.2 Current
output
Next, the screens for programming the current loop appear. The flow rate units will be
those selected previously.
12
3.2
Configuration of the rest of the parameters.
Touching the two keys (↑) & (
□
) at the same time the configuration sequence for the
display and output parameters is accessed.
3.2.1 Decimals
In the first screen the number of decimals for the flow rate indication is configured.
For this, touching the key (↑) the flashing digit will be increased. When the value of 2 is
reached, on the next increment the digit will go to zero. When we have the required factor
on the screen, touching the two keys (↑) & (←) at the same time, the data will be stored in
memory.