Super Systems 7EK 31080 User Manual
Page 27
23
The sensor leads break can be signalled as:
- for TC/mV input
:
OVERRANGE or UNDERRANGE
selected by a solder jumper
- for RTD input
:
OVERRANGE
- for mA/V input
:
UNDERRANGE
Note: On the mA/V input the leads break can be detected
only when the range selected has a zero elevation (4/20
mA or 1/5 V or 2/10 V)
On RTD input a special test is provided to signal
OVERRANGE when input resistance is less than 15 ohm
(Short circuit sensor detection).
ERROR MESSAGES
The instrument performs some self-diagnostic tests.
When an error is detected, the instrument shows on the
lower display the “Err” indication, while the upper display
shows the code of the detected error.
ERROR LIST
SEr
Serial interface parameter error.
100
Write EEPROM error.
200
Attempt to write on protected register.
201 - 2xx
Configuration parameter error. The two
less significant digits show the number of
the wrong parameter (ex. 209 Err shows
an Error on P9 parameter).
301
RTD input calibration error.
305
TC and mV input calibration error.
307
RJ input calibration error.
310
Current transformer input calibration error.
311
Current input (20 mA) calibration error.
312
5 V input calibration error.
313
10 V input calibration error.
400
Operative parameters error
500
Auto-zero error
502
RJ error
510
Error during calibration procedure
NOTES:
1) When a configuration parameter error is detected, it is
sufficient to repeat the configuration procedure of the
specific parameter.
2) If error 400 is detected, push contemporarily the
t
and
s
pushbuttons for loading the default parameters
then repeat run time parameter setting.
3) For all the other errors, contact your supplier.
31080-1-00.p65
3/24/00, 11:58 AM
23