Super Systems 7EK 31080 User Manual
Page 10
6
A
A
B
MAIN INPUT SELECTION
If an input different from TC-RTD input (standard) is desired,
remove the instrument from its case and set J1 according to
the following table.
INPUT
J1
TYPE
1-2
3-4
5-6
7-8
9-10
TC-RTD
open
close
open
open
open
60 mV
open
close
open
open
open
5 V
close
open
close
open
open
10 V
open
open
close
open
open
20 mA
open
open
open
close
close
NOTE: the jumper not used can be placed on pin 7-9
PRELIMINARY HARDWARE SETTINGS
How to remove the instrument from its case
1) Switch off the instrument.
2) Push gently the lock A on the right.
3) While the lock A is maintained out, slide out the right
side of the instrument.
4) Push gently the lock B on the left.
5) While the lock B is maintained out, slide out the
instrument.
Fig. 11
1 3 5 7 9
2 4 6 8 10
J1
31080-1-00.p65
3/24/00, 11:58 AM
6
- Bazooka Probe (10 pages)
- Gold Probe (16 pages)
- HP2000 With 9100 Controller (10 pages)
- HP15 (23 pages)
- SuperOX (14 pages)
- PGA3000 (16 pages)
- PGA3500 (26 pages)
- e-TRIM (27 pages)
- 9120 with TS (80 pages)
- MGA6000 (42 pages)
- DP2000 (17 pages)
- DPC3500 (5 pages)
- MGA6010 (54 pages)
- DPC2530 (17 pages)
- Simple Dew (18 pages)
- DPL4000 (16 pages)
- H2 Sensor (17 pages)
- Hydrogen Nitrider Analyzer (12 pages)
- PH2 (19 pages)
- AC20 Quick Start (5 pages)
- XGA Viewer (46 pages)
- AC20 RS485 Modbus (62 pages)
- AC20 (114 pages)
- 20Q Calibration Manual (9 pages)
- CAT-100 (51 pages)
- 7EK 31082 Calibration Manual (5 pages)
- 7EK 31081 (36 pages)
- 7EK 31082 (34 pages)
- 20PQ (170 pages)
- X5 Calibration Manual (1 page)
- 20Q (126 pages)
- 7SL (36 pages)
- X5 (74 pages)
- 9000 Series (15 pages)
- 3L Series (54 pages)
- 9015 Series (11 pages)
- 3 Series (92 pages)
- 9010 Series (66 pages)
- 9210 Series (65 pages)
- 9130 Series (174 pages)
- PC Configurator 2 Quick Start (15 pages)
- 9100 RPS (10 pages)
- 9125 Series (235 pages)
- 9120 RPS (11 pages)