SSi Manual #4563 Rev. D Page 4 Multi-Gas Model 6000
MGA6000 Operating Instructions
Introduction
The Model MGA6000 is a continuous Multi-Gas Analyzer using NDIR (Non-Dispersive Infra-Red) technology. It is
capable of measuring between one and four gases, depending on how it is configured at the factory prior to shipment.
The unit is designed and manufactured for the atmosphere heat treating industry.
Possible Configurations
CO range:
0.00 to 30.00 % *
CO
2
range:
0.000 to 2.000 %
CO
2
High range:
0.0 to 20.00 %
CH
4
range:
0.00 to 15.00 % *
O
2
range:
0.1 to 25.0%
Hydrogen:
0.0 to 100.0%
Dew point
0 to 80 F
* Note: The CO and CH4 sensors have been calibrated to be most accurate within the ranges as shown
above, however both sensors are capable of measuring gas concentrations of up to 100.00%.
Specifications
Sampling method:
Extraction by internal pump (when necessary)
Accuracy and repeatability:
± 1% of full scale
Flow Meter:
0 to 2 SCFH, mounted on enclosure front and also displayed on-
screen
Pump Operation:
On/Off/Auto Sample
AC Power:
90 to 230 VAC, 50 to 60 Hz, 60 Watts
Communications:
Ethernet,
RS485
4-20 mA Outputs:
Four (4) user-defined
Digital Inputs:
Two (2) for stopping pump and inhibiting COF/PF adjustment
Relays:
One (1) for alarm or control
Operating Temperature: 32
° to 122° F (0° to 50° C)
Dimensions:
Approx. 16” x 14” x 8”
Weight:
Approx.
20
lbs.