Super Systems MGA6000 User Manual
Page 29

SSi Manual #4563 Rev. D Page 29 Multi-Gas Model 6000
65 flow
span
value
* Register 66 not used
67 Dualpro
process
factor
register
68
LED D2 ON time in millisecs
69
LED D2 OFF time in millisecs
70 Display
polarity
71
Display contrast
72
Backlight brightness
73
Backlight on time; 0 = always ON
74
Web change enable; 0 = disable, 1 = enable
75
Calibration
stage
76
Pump ON delay
77 Sample
OFF
delay
78
Adjust/control update interval
79
sample ON delay time
80
H2 zero gas % times 100 (xx.xx)
81
H2 span gas % times 100 (xx.xx)
82
hydrogen cell calibration request; 1 = zero, 2 = span
83
H2 cell in % times 100 (xx.xx)
84
Help
pointer
85
Sum of Active Cell start up timers
86
Backlight
brightness
output
87 to 95
slave communications status
96
PIC processor communications status (96 - 99)
100
local cooler ON set point
101
local cooler OFF set point
102
auto calibration and/or sequencing
bit 0 = auto cal; 0 = OFF, 1 = ON
bit 1 = sequencing; 0 = OFF, 1 = ON
bit 2 = seq mode; 0 = normal; 1 = specific
bits 4 - 7 Not Used
bits 8 - 13 CH active bit map
bits 14 - 15 Not Used
103
Auto cal interval in minutes 0 = OFF
104
Auto Zero interval in minutes 0 = OFF
105
Auto data display time at end of seq
106
purge time before cal or zero
* Registers 107 to 109 not used
* Selected gas types:
0 = none
1 to 4 = PIC 0 to 3 SPI IR cells respectively
5 = H2 cell via RS232
6 = Lambda probe via RS-485
7 = PIC 2 A/D1 Oxygen cell
8 = PIC 2 A/D3 Ext thermister
9 = PIC 1 A/D3 Ext 0 to 5 volt
10 = PIC 3 A/D0 Ext 0 to 1 volt
11 = PIC 3 A/D1 Ext 0 to 1 volt
12 = PIC 3 A/D3 Ext 0 to 5 volt
Selected Calculation types:
0 = none
1 = IR % carbon (needs CO and CO2 and optionally CH4)
2 = Gas ratio (Gas_1_val/Gas_2_val)