879
If none, zero inhibits datalogging
880
Auto mount return code
881
Number of FAT devices available
882
MAX number of possible partitions
883
Partitions mounted (883 - 890)
891
File open return code
892
Number of log partition, 1 = A:, 5 = E:
893
return code of UDPDL_init()
* Registers 894 to 899 not used
900
product ID code
901
reset logging
902
Dynamic C compiler version in hex
903
long date/time stamp in secs from
904
midnight Jan 1, 1980.
905
Hack attempts counter
906
Web access code, level 1
907
Web access code, level 2
908
session ID passed to Java Applet
909
Set factory defaults control, 23205 (0x5aa5) to set
910
Force user block write; 0xa5 (165) = write w/o conf,
0xa9 (169) = write with config, 222 = write user setups, 444 = read user
setups
* Registers 911 to 913 not used
914
4 locations for IP address (914 - 917)
918
4 locations for netmask (918 - 921)
922
4 locations for gateway (922 - 925)
* Registers 926 to 929 not used
930
10 locations for port states (930 - 939)
940
auto calibration interval timer
941
auto zero interval timer
942
auto cal/zero request; bit 0 = zero, bit 1 = span
943
auto cal/zero sequence
944
sequence timer
945
auto cal/zero status
946
Flag to initiate log of Auto cal
947
Number of entries in Auto cal log file
* Registers 948 to 949 not used
950
Auto
sequence
pointer
951
Auto
sequence
stage
952
Auto
sequence
timer
* Registers 953 to 995 not used
996
web page writes here as update indicator (996 - 999)
* Events IN/Output file_1[EVT_base][] offsets
1603
event output setpoint
1605
event output actual
1606
event output new setpoint
1698
events setpoint write reg
* Analog output file_1[DACbase][] offsets
1700
copy of com_stat + DACbase -1
1701
board address (lo Byte)
1701