Sensoray 526 User Manual
Page 14

14
−
−
⋅
=
0
0
C
C
C
C
V
V
ref
meas
ref
meas
,
where
ref
V
- the actual value of the on-board 10V reference (from the EEPROM);
meas
C
- ADC reading corresponding to the measured voltage;
ref
C
- ADC reading corresponding to the on-board +10V reference;
0
C
- ADC reading corresponding to the on-board 0V reference.
The calibration constant is stored in the
double
floating point format (8 bytes per value). See
the
Calibration EEPROM
section for the details.