beautypg.com

Primetest 200, Operating instructions 20 – Seaward 200 PrimeTest User Manual

Page 30

background image

Point to Point Insulation

Intrinsic Error or

Reference Conditions or

Code

Max Error

Influence Quantity

Rated Operating Conditions

Seen

Intrinsic Error

Reference Condions

A

5%

Position

Reference Positions ±90°

E

1

N/A

Supply Voltage

At the Limits Specified

E

2

1.6%

by the Manufacturer

Temperature

0°C and 35°C

E

3

1%

Absorbed Current of

Maximum Absorbed Current

Device on Test

of the Device on Test

E

4

N/A

Low-frequency

Low-frequency

Magnetic Field

Magnetic field 0.4 kA/m

E

5

N/A

Not Used

Not Used

E

6

N/A

Capacitance

Cmax = 1uF

E

7

2%

Waveform

R.M.S. CR <= 2

E

8

N/A

Operating Error

B=\A\+1.15

E

1

2

+E

2

2

+E

3

2

+E

4

2

+E

5

2

+E

6

2

+E

7

2

+E

8

2

7.17%

PrimeTest 200

Operating Instructions

20

Differential Leakage

Intrinsic Error or

Reference Conditions or

Code

Max Error

Influence Quantity

Rated Operating Conditions

Seen

Intrinsic Error

Reference Condions

A

5%

Position

Reference Positions ±90°

E

1

N/A

Supply Voltage

At the Limits Specified

E

2

0.56%

by the Manufacturer

Temperature

0°C and 35°C

E

3

0.81%

Absorbed Current of

Maximum Absorbed Current

Device on Test

of the Device on Test

E

4

1.5%

Low-frequency

Low-frequency

Magnetic Field

Magnetic field 0.4 kA/m

E

5

N/A

Not Used

Not Used

E

6

N/A

Capacitance

Cmax = 1uF

E

7

N/A

Waveform

R.M.S. CR <= 2

E

8

2.5%

Operating Error

B=\A\+1.15

E

1

2

+E

2

2

+E

3

2

+E

4

2

+E

5

2

+E

6

2

+E

7

2

+E

8

2

8.54%

Touch Leakage

Intrinsic Error or

Reference Conditions or

Code

Max Error

Influence Quantity

Rated Operating Conditions

Seen

Intrinsic Error

Reference Condions

A

10%

Position

Reference Positions ±90°

E

1

N/A

Supply Voltage

At the Limits Specified

E

2

2.7%

by the Manufacturer

Temperature

0°C and 35°C

E

3

1.04%

Absorbed Current of

Maximum Absorbed Current

Device on Test

of the Device on Test

E

4

0.87%

Low-frequency

Low-frequency

Magnetic Field

Magnetic field 0.4 kA/m

E

5

N/A

Not Used

Not Used

E

6

N/A

Capacitance

Cmax = 1uF

E

7

N/A

Waveform

R.M.S. CR <= 2

E

8

0.79%

Operating Error

B=\A\+1.15

E

1

2

+E

2

2

+E

3

2

+E

4

2

+E

5

2

+E

6

2

+E

7

2

+E

8

2

13.59%