7.0

Electrical Specification

Appendix 1 specifies the operating error for all tests and functions of the
PrimeTest 200.

7.1

Earth continuity measuring range (rotary switch position , and IEC)

Measuring Range

Resolution

Accuracy

Overload Protection

19.99

0.01

±5% +2 digits

300 V AC/DC

Analogue bargraph display

0 to 20

Bargraph accuracy

10%

Open circuit voltage

>4VDC

Test current

>200mA (0-2

)

Input protection

0.5A (600V) quick blow fuse

An input voltage of

30 V AC/DC, detected between the + and - input measuring

terminals, disables the earth continuity measuring circuit. A repetitive audible
alarm is activated and a flashing high voltage symbol is displayed on the LCD.

The maximum number of continuity tests, with a duty cycle specified by
BS EN 61557-4 Section 6.6, is 1900 repetitions.

7.2

Insulation resistance measuring range (rotary switch position , , IEC
and M

.

Measuring Range

Resolution

Accuracy

Overload Protection

2M

0.01M

±2% +2 digits

300 V AC/DC

20M

0.1M

±2% +2 digits

300 V AC/DC

200M

1M

±5% +2 digits

300 V AC/DC

Analogue bargraph display

0 to 200M

and infinity

Bargraph accuracy

10%

Test voltage

>500VDC at 1mA, <750VDC for o/c

Test voltage accuracy

+20%, -0%

Test current

<2mA for s/c

Input protection

0.5A (600 V) quick blow fuse.

An input voltage of

30 V AC/DC, detected between the + and - input measuring

terminals, disables the insulation resistance measuring circuit. A repetitive
audible alarm is activated and a flashing high voltage symbol is displayed on
the LCD.

The maximum number of insulation tests, with a duty cycle specified by
BS EN 61557-2 Section 6.7, is 2000 repetitions.

7.3

Resistance measuring range

Measuring Range

Resolution

Accuracy

Overload Protection

19.99

0.01

±5% +2 digits

300 V AC/DC

1999

1

±5% +2 digits

300 V AC/DC

20.00K

10

±5% +2 digits

300 V AC/DC

Analogue bargraph display

0 to 20K

and infinity

Bargraph accuracy

10%

Test current

>200mA (0-2R)

Open circuit voltage

> 4 V DC

Input protection

0.5A (600 V) quick blow fuse.

Zero compensation of the test leads up to 10

. The internal buzzer sounds

when the measured resistance <10

.

An input voltage of

30 V AC/DC, detected between the + and - input measuring

terminals, disables the resistance measuring circuit. A repetitive audible alarm is
activated and a flashing high voltage symbol is displayed on the LCD.