Primetest 200, Operating instructions 19 – Seaward 200 PrimeTest User Manual
Page 29
![background image](https://www.manualsdir.com/files/804190/content/doc029.png)
IEC Insulation
Intrinsic Error or
Reference Conditions or
Code
Max Error
Influence Quantity
Rated Operating Conditions
Seen
Intrinsic Error
Reference Condions
A
5%
Position
Reference Positions ±90°
E
1
N/A
Supply Voltage
At the Limits Specified
E
2
1%
by the Manufacturer
Temperature
0°C and 35°C
E
3
1%
Absorbed Current of
Maximum Absorbed Current
Device on Test
of the Device on Test
E
4
N/A
Low-frequency
Low-frequency
Magnetic Field
Magnetic field 0.4 kA/m
E
5
N/A
Not Used
Not Used
E
6
N/A
Capacitance
Cmax = 1uF
E
7
N/A
Waveform
R.M.S. CR <= 2
E
8
N/A
Operating Error
B=\A\+1.15
√
E
1
2
+E
2
2
+E
3
2
+E
4
2
+E
5
2
+E
6
2
+E
7
2
+E
8
2
6.63%
PrimeTest 200
Operating Instructions
19
Class 2 Insulation
Intrinsic Error or
Reference Conditions or
Code
Max Error
Influence Quantity
Rated Operating Conditions
Seen
Intrinsic Error
Reference Condions
A
5%
Position
Reference Positions ±90°
E
1
N/A
Supply Voltage
At the Limits Specified
E
2
0.53%
by the Manufacturer
Temperature
0°C and 35°C
E
3
1%
Absorbed Current of
Maximum Absorbed Current
Device on Test
of the Device on Test
E
4
N/A
Low-frequency
Low-frequency
Magnetic Field
Magnetic field 0.4 kA/m
E
5
N/A
Not Used
Not Used
E
6
N/A
Capacitance
Cmax = 1uF
E
7
N/A
Waveform
R.M.S. CR <= 2
E
8
N/A
Operating Error
B=\A\+1.15
√
E
1
2
+E
2
2
+E
3
2
+E
4
2
+E
5
2
+E
6
2
+E
7
2
+E
8
2
6.30%
Point to Point Resistance
Intrinsic Error or
Reference Conditions or
Code
Max Error
Influence Quantity
Rated Operating Conditions
Seen
Intrinsic Error
Reference Condions
A
5%
Position
Reference Positions ±90°
E
1
N/A
Supply Voltage
At the Limits Specified
E
2
0.6%
by the Manufacturer
Temperature
0°C and 35°C
E
3
1.4%
Absorbed Current of
Maximum Absorbed Current
Device on Test
of the Device on Test
E
4
N/A
Low-frequency
Low-frequency
Magnetic Field
Magnetic field 0.4 kA/m
E
5
N/A
Not Used
Not Used
E
6
N/A
Capacitance
Cmax = 1uF
E
7
N/A
Waveform
R.M.S. CR <= 2
E
8
N/A
Operating Error
B=\A\+1.15
√
E
1
2
+E
2
2
+E
3
2
+E
4
2
+E
5
2
+E
6
2
+E
7
2
+E
8
2
6.75%