Primetest 200 – Seaward 200 PrimeTest User Manual

Page 16

background image

PrimeTest 200

Operating Instructions

6

7.4

Voltage measuring range

Measuring Range

Resolution

Accuracy

Overload Protection

250 V AC/DC

1V

±2% +2 digits

300 V AC/DC

Frequency Range

up to 400Hz

Analogue bargraph display

0 to 250 V

Bargraph accuracy

10%

Input protection

>20M

An input voltage of

30 V AC/DC, detected between the + and - input measuring

terminals, disables the voltage measurement function. A repetitive audible alarm
is activated and a flashing high voltage symbol is displayed on the LCD.

7.5

IEC lead test

The IEC lead test is a sequence of earth continuity measurement (section 7.1)
- insulation resistance measurement (section 7.2) - Live/Neutral checks for o/c,
s/c and crossed conductors in the IEC lead.

Test voltage

9 VDC

Test duration

10 seconds

7.6

Temperature measuring range °C

Measuring Range

Resolution

Accuracy

Overload Protection

-50°C to +400°C

0.1°C

±5% +2digits

300 V AC/DC

An input voltage of

30 V AC/DC, detected between the + and - input measuring

terminals, disables the temperature measurement function. A repetitive audible
alarm is activated and a flashing high voltage symbol is displayed on the LCD.

7.7

Touch current measurement

Measuring Range

Resolution

Accuracy

Overload Protection

3.5mA

0.01mA

±10% + 2 digits (5° - 40°)

N/A

±15% + 2 digits (0° - 5°)

Test voltage

mains supply @ 230V - 15%/+10% @ 10A max

Frequency

50Hz ±1%

Test duration

2 seconds minimum

Frequency response

IEC 61010-1 Annex A.1

An input voltage of

30 V AC/DC, detected between the + and - input measuring

terminals, disables the touch current test function. A repetitive audible alarm is
activated and a flashing high voltage symbol is displayed on the LCD.

The test is also disabled when phase-neutral (PN) or phase-earth (PE) voltage
is not at mains potential or a voltage of 40 V AC or greater exists between
neutral-earth (NE).

7.8

Differential current measurement

Measuring Range

Resolution

Accuracy

Overload Protection

19.99m A

0.01mA

±5% + 2 digits

N/A

Test voltage

mains supply @ 230V - 15%/+10% @ 10A max

Frequency

50Hz ±1%

Test duration

2 seconds minimum

Frequency response

IEC 61010-1 Annex A.1 (from 40Hz)

An input voltage of

30 V AC/DC, detected between the + and - input measuring

terminals, disables the differential current test function. A repetitive audible
alarm is activated and a flashing high voltage symbol is displayed on the LCD.

The test is also disabled when phase-neutral (PN) or phase-earth (PE) voltage
is not at mains potential or a voltage of 40 V AC or greater exists between
neutral-earth (NE).