CAUTION: Use of controls, adjustments or performance of procedures other
than those specified in this Operation Manual may result in hazardous
radiation exposure.
Follow the instructions within this manual carefully to ensure the safe operation of your
laser. At all times during laser operation, maintenance, or servicing, avoid unnecessary
exposure to laser or collateral radiation that exceeds the accessible emission limits listed in
"Performance Standards for Laser Products," United States Code of Federal Regulations,
21 CFR 1040.10(d). This information is also available in EN60825-1:1994, Section 8.2,
titled “Measurements of Laser Radiation for Determining Classification.”
Preventative Maintenance for Safety: Preventative maintenance is required to ensure the
laser remains in compliance with Center for Devices and Radiological Health (CDRH)
Regulations and European Norm (EN) requirements. This laser product complies with Title
21 of the United States Code of Federal Regulations, Chapter 1, Subchapter J, Parts
1040.10, as applicable, and with EN60825-1:1994, Part 1 for a Class 4 laser, as applicable.
To maintain compliance, verify the operation of all features listed below, either annually or
whenever the product has been subjected to adverse environmental conditions, which may
have affected these features and functions.
o
Verify that removing the remote interlock connector prevents laser operation. This
connector is located on the end cap of the laser.
o
Verify that the laser will operate only with the key switch in the ON position and the
key can only be removed when the switch is in the OFF position.
o
Verify that a time delay exists between turning on the Key Switch and the start of
laser firing. The time delay must be long enough to allow action to be taken to avoid
exposure to laser radiation.
ELECTRICAL SAFETY
CAUTION: Both the Laser Optics Assembly and Laser Power Supply contain
electrical circuits operating at lethal voltage and current levels. Always
unplug the system Mains connection and wait at least one (1) minute to allow
capacitors to discharge before servicing any part of the laser system.
Consult with the Quantum Composers if repair of the laser electronics is required. Only those
trained in high voltage, high current electronics, and who understand the laser circuitry, should be
allowed to service and repair the laser electronics. If any such action is required, it is
recommended that you contact the Quantum Composers for details.
Sources of Laser Safety Standards
"Safe Use of Lasers" (Z136.1)
American National Standards Institute (ANSI)
11th West 42nd Street
New York, NY 10036 USA
Phone: (212) 642-4900
"A Guide for Control of Laser Hazards"