Quantum Composers Jewel Laser User Manual
Page 24

Page 24 Jewel Manual v1.1 | Quantum Composers
SC?
System shot count, since factory build.
SS?
System Status – Query Only. Returns the current system state.
Value is a 16 bit decimal value with each bit position
corresponding to a system state (See Table below, bits 0-15).
15
14
13
12
11
10
9
8
Spare
Spare
Spare
Spare
Ready to
Fire
Ready
to
Enable
Power
Failure
Electrical
Over
Temp
(FET)
7
6
5
4
3
2
1
0
Resonator
Over
Temp
External
Interlock
Shutter
Closed
Q-Switch
External
Trigger
Diode
External
Trigger
Q-
Switch
Enabled
Laser
Active
Laser
Enabled
SV #
SV?
Save settings to user bin 1-6.
SY #
SY?
Sync Output Mode. 0 = off, 1 = LD pulse, 2 = QSW pulse, 3 =
LD/QSW.
TA?
Resonator thermistor A temp. In degrees C.
TB?
Resonator thermistor B temp. In degrees C.
UC #
UC?
User shot count. Saved shot count. Can be cleared on command
by issuing a 0 after command. Can also be preset by sending any
number up to 32 bit.
Harmonic Module Heater Commands (Note: Applicable if Installed)
H1?
Heater 1 installed (0 = No, 1 = Yes)
H1:SP?
Heater 1 setpoint. Target heater setting, in degrees C.
H1:ST?
Heater 1 Status byte (See Table Below).
H1:TP?
Heater 1 temperature. Active temperature of the heater in degrees
C.
H1:VN?
Heater 1 controller firmware version.
H2?
Heater 2 installed (0 = No, 1 = Yes)
H2:SP?
Heater 2 setpoint. Target heater setting, in degrees C.
H2:ST?
Heater 2 Status byte (See Table Below).
H2:TP?
Heater 2 temperature. Active temperature of the heater in degrees
C.
H2:VN?
Heater 2 controller firmware version.
7
6
5
4
3
2
1
0
spare
spare
spare
Temp
reading
out of
range
Temp
too high
At
Target
Temp
Temperature
settling
Temp
Control
Enabled