Laser status connectors, Initial power on, Power off – Quantum Composers Jewel Laser User Manual
Page 20

Page 20 Jewel Manual v1.1 | Quantum Composers
LASER STATUS CONNECTORS
1. Power LED – Illuminated when Key Switch is ON
2. Fault – Illuminated when laser fault exists. Laser will not fire when FAULT LED is
illuminated.
3. Laser Emission – LED is illuminated any time the laser is enabled (ready to fire) or when
actively lasing.
INITIAL POWER ON
Power-On Checklist
Verify cable connections are correct.
Apply power to laser power supply.
Rotate keyswitch clockwise to turn on the Jewel Laser.
Open Jewel Control Software.
The laser will be ready to operate after approximately 2 minutes. When harmonic module is
attached, best laser performance will require 10 minutes of warm-up time.
POWER OFF
Power-Off Checklist
Stop laser Fire, via software.
Close Jewel Control Software.
Power down Jewel Laser.