Laser Operation Commands
BC #
BC?
Burst Count. Sets the number of laser shots to be fired when firing
mode is set to burst. Parameters are from 1-1000. Default is 10.
A query will return the currently set burst count.
EC #
EC?
Echo. Parameters: 0=echo characters off (Default: 0), 1=echo
characters on. EC? returns echo state.
EN #
EN?
Enable. This will enable or disable the laser. The laser must first
be enabled before firing. 0 = Disabled, 1 = Enabled. (Default: 0).
A query returns the current enable state.
FL #
FL?
Fire Laser. Starts the laser pulsing.
0 = Stop Firing Laser, place laser in idle state, shutter will be
closed.
1 = Fire laser, shutter is closed.
2 = Fire laser, shutter is open.
A query returns the current firing state.
FT?
FT:MAX?
FET temperature in degrees C.
FV?
FET Voltage in Volts.
ID?
System ID – QC, model #, serial # (5 digits),version# (x.x) (ex.
QC,JEWEL LASER,00101,0.1). Query only.
IM?
Diode current measurement, in Amps.
LS?
Latched Status. Returns the latched system status. Cleared
upon read.
PE #.#
PE?
PE:MIN?
PE:MAX?
Sets pulse period in seconds.
PM #
PM?
Mode. Sets the laser firing mode, 0 = continuous, 1 = single shot,
2 = burst. (Default: 0). A query returns the currently set mode.
QD #.#
QD?
QD:MIN?
QD:MAX?
Sets Q-Switch delay (from diode pulse rising edge) in seconds.
QE #
QE?
Q-Switch Enable. Enables normal Qswitch operation. 0 –
Disables normal operation, 1 – Enables normal operation.
QT #
QT?
Q-Switch trigger mode. 0 = internal, 1 = external, default = internal
QW #
QW?
Q-Switch Wait. The number of diode pulses that the q-switch will
wait before activating. A zero will disable the wait. (Range 0 –
65535)
RC #
RC?
Recall settings from user bin 1-6. 0 = recall factory defaults.
RR #.#
RR?
RR:MIN?
RR:MAX?
Repetition Rate. This sets the rate at which the laser will fire at.
Values are from 1-40 Hz. (Default: 10 Hz). A query will return the
current rate.
RS
Reset. Resets the system.