Quantum Composers 9730 Series User Manual
Page 41

10-41
Some customers have started using the 9730 device in high throughput production lines and have
requested that this safety feature be removed. Quantum Composers has added a bypass
feature to accommodate these customers under the condition that usage of this bypass
function is an acceptance agreement by the user that they are knowingly and willingly
bypassing a safety feature for the 9730 and accept any liability associated with bypassing
such feature.
Bypass Instructions
The safety feature can by bypassed by sending a command after the unit has been powered on.
This command can be sent along with any other initialization commands that may already being
sent on startup. This command is a static command such that when the unit is powered off, it will
return to the default safe operation until the initialization command is sent again. The commands
are as follows:
:SPUL:ARMD 0
(This will disable the feature of disabling the channels and not
automatically re-enabling them).
:SPUL:ARMD 1
(This will restore the safety feature of disabling the channels and not
automatically re-enabling them)
:SPUL:ARMD?
(This is a query command to read the current state of the bypass
command)