Single shot mode, System burst mode function, System duty cycle function – Quantum Composers 9250 Series User Manual
Page 26

9250 Manual Version 1.2 | Quantum Composers
26
Single Shot Mode
Pressing the unit’s Run/Stop button generates a single pulse with every press. Set the
following parameters in the 9250 application:
Within the System section:
o Select Single Shot mode.
Within the Channel section:
o Set channel mode to Normal.
o Enter the required pulse width and delay. Repeat for each output channel.
o
Enable the corresponding channels by clicking “Enabled” in the 9250 application.
Pressing the unit’s Run/Stop button will now generate one pulse out of every enabled channel.
System Burst Mode Function
The Run/Stop button
generates a stream of ‘N’ T
0
pulses, where the ‘N’ is specified by the
Burst parameter. The rate is specified by the Period parameter. Pressing the
Run/Stop
button
while the burst is in process will stop the output. After the burst has been completed, Pressing
the
Run/Stop
button will generate another burst of pulses. To generate a burst of pulses set
the following parameters in the 9250 application:
Within the System section:
o Select mode to be Burst.
o Set the Burst Count parameter field to produce the number of pulses desired
o Set the desired Period.
Within the Channel section:
o Set channel mode to Normal.
o Enter the required pulse width and delay. Repeat for each output channel.
o
Enable the corresponding channels by clicking “Enabled” in the 9250 application.
Pressing the unit’s Run/Stop button will now generate the predefined burst of pulses out of
every enabled channel.
System Duty Cycle Function
The Run/Stop button starts a continuous stream of T
0
pulses, which repeats
for ‘N’ pulses On
and ‘M’ pulses Off, where ‘N’ and ‘M’ are specified by the On/Off parameters respectively. The
rate at which the pulses are generated is controlled by the Period parameter. To generate a
stream of pulses which will
repeat for ‘N’ pulses On and ‘M’ pulses, set the following
parameters in the 9250 application:
Within the System section:
o Set the mode to Duty Cycle.
o Set the On parameter to number of pulses to produce during th
e on cycle (‘N’).
o Set the Off parameter to
number of pulses to suppress during the off cycle (‘M’).
o Set the desired Period.