beautypg.com

Analog voltage output – Measurement Computing USB-2416-4AO User Manual

Page 37

background image

USB-2416-4AO User's Guide

Specifications

37

Analog voltage output

Unused VDACx output channels should be left disconnected.

The USB-2416-4AO output voltage level defaults to 0 V whenever the host PC is reset, shut down or
suspended, or if a reset command is issued to the device.

The duration of the output transient depends highly on the enumeration process of the host PC. Typically, the
output of the USB-2416-4AO is stable after two seconds.

Table 20. Analog voltage output specifications

Parameter

Conditions

Specifications

Digital-to-analog
converter

DAC8555

Number of channels

4

Resolution

16 bits

Output ranges

Calibrated

±10 V

Uncalibrated

±10.05 V, software configurable

Output transient

Host PC is reset, powered on, suspended or a reset
command is issued to device (refer to the output
transient discussion above)

Duration: 2 s
Amplitude: 2 V p-p

Initial power on

Duration: 50 ms
Amplitude: 5 V peak

Differential
non-linearity

±0.25 LSB typical
±1 LSB maximum

Output current

VDACx pins

±3.5 mA maximum

Output short-circuit
protection

VDACx connected to AGND

Unlimited duration

Output coupling

DC

VDACx readback

Each VDACx output can be
independently measured by the onboard
A/D converter.
Software selectable

Power on and reset
state

DACs cleared to zero-scale: 0 V,
±50 mV

Output noise

30 µVrms

Settling time

To rated accuracy, 10 V step

45 µs

Slew rate

1.0 V/µs

Throughput

Single-channel

1000 S/s maximum, system-dependent

Multi-channel

1000 S/s / #ch maximum, system-
dependent

Table 21. Calibrated absolute accuracy specifications

Range

Accuracy (±LSB)

±10 V

16.0

Table 22. Calibrated absolute accuracy components specifications

Range

% of reading

Offset (±mV)

Temp drift (%/°C)

Absolute accuracy at FS (±mV)

±10 V

±0.0183

1.831

0.00055

3.661

Table 23. Relative accuracy specifications

Range

Relative accuracy (±LSB)

±10 V

4.0 typical