beautypg.com

Measurement Computing USB-2416-4AO User Manual

Page 24

background image

USB-2416-4AO User's Guide

Specifications

24

The USB-2416-4AO GND and DGND pins are isolated from earth ground. You can connect thermocouple
sensors to voltages referenced to earth ground as long as isolation between the GND/DGND pins and earth
ground is maintained.

Table 4. J type thermocouple accuracy specifications, including CJC measurement error.

All specifications are (±).

A/D data rate

Sensor temperature range

Accuracy error
maximum

Accuracy error
typical

Tempco

(°C/°C)

3750 S/s

-210 °C

2.572 °C

1.416 °C

0.022

0 °C

0.935 °C

0.469 °C

1200 °C

1.869 °C

1.456 °C

2000 S/s

-210 °C

2.572 °C

1.416 °C

0.022

0 °C

0.935 °C

0.469 °C

1200 °C

1.869 °C

1.456 °C

1000 S/s

-210 °C

2.572 °C

1.416 °C

0.022

0 °C

0.935 °C

0.469 °C

1200 °C

1.869 °C

1.456 °C

500 S/s

-210 °C

2.442 °C

1.334 °C

0.022

0 °C

0.881 °C

0.415 °C

1200 °C

1.821 °C

1.408 °C

100 S/s

-210 °C

2.442 °C

1.334 °C

0.022

0 °C

0.881 °C

0.415 °C

1200 °C

1.821 °C

1.408 °C

60 S/s

-210 °C

2.442 °C

1.334 °C

0.022

0 °C

0.881 °C

0.415 °C

1200 °C

1.821 °C

1.408 °C

50 S/s

-210 °C

2.442 °C

1.334 °C

0.022

0 °C

0.881 °C

0.415 °C

1200 °C

1.821 °C

1.408 °C

25 S/s

-210 °C

2.442 °C

1.334 °C

0.022

0 °C

0.881 °C

0.415 °C

1200 °C

1.821 °C

1.408 °C

10 S/s

-210 °C

2.442 °C

1.334 °C

0.022

0 °C

0.881 °C

0.415 °C

1200 °C

1.821 °C

1.408 °C

5 S/s

-210 °C

2.442 °C

1.334 °C

0.022

0 °C

0.881 °C

0.415 °C

1200 °C

1.821 °C

1.408 °C

2.5 S/s

-210 °C

2.442 °C

1.334 °C

0.022

0 °C

0.881 °C

0.415 °C

1200 °C

1.821 °C

1.408 °C

To achieve the thermocouple accuracies listed above, the USB-2416-4AO should be warmed up for 45 minutes
after the initial power on. The accuracies listed above are only guaranteed if the USB-2416-4AO is housed in its
plastic enclosure.