beautypg.com

Measurement Computing USB-2416-4AO User Manual

Page 26

background image

USB-2416-4AO User's Guide

Specifications

26

Table 6. N type thermocouple accuracy specifications, including CJC measurement error.

All specifications are (±).

A/D data

rate

Sensor
temperature range

Accuracy error maximum

Accuracy error
typical

Tempco
(°C/°C)

3750 S/s

-200 °C

3.480°C

2.030 °C

0.029

0 °C

1.201 °C

0.659 °C

1300 °C

1.991 °C

1.600 °C

2000 S/s

-200 °C

3.480°C

2.030 °C

0.029

0 °C

1.201 °C

0.659 °C

1300 °C

1.991 °C

1.600 °C

1000 S/s

-200 °C

3.480°C

2.030 °C

0.029

0 °C

1.201 °C

0.659 °C

1300 °C

1.991 °C

1.600 °C

500 S/s

-200 °C

3.196°C

1.750 °C

0.029

0 °C

1.096 °C

0.553 °C

1300 °C

1.915 °C

1.524 °C

100 S/s

-200 °C

3.196°C

1.750 °C

0.029

0 °C

1.096 °C

0.553 °C

1300 °C

1.915 °C

1.524 °C

60 S/s

-200 °C

3.196°C

1.750 °C

0.029

0 °C

1.096 °C

0.553 °C

1300 °C

1.915 °C

1.524 °C

50 S/s

-200 °C

3.196°C

1.750 °C

0.029

0 °C

1.096 °C

0.553 °C

1300 °C

1.915 °C

1.524 °C

25 S/s

-200 °C

3.196°C

1.750 °C

0.029

0 °C

1.096 °C

0.553 °C

1300 °C

1.915 °C

1.524 °C

10 S/s

-200 °C

3.196°C

1.750 °C

0.029

0 °C

1.096 °C

0.553 °C

1300 °C

1.915 °C

1.524 °C

5 S/s

-200 °C

3.196°C

1.750 °C

0.029

0 °C

1.096 °C

0.553 °C

1300 °C

1.915 °C

1.524 °C

2.5 S/s

-200 °C

3.196°C

1.750 °C

0.029

0 °C

1.096 °C

0.553 °C

1300 °C

1.915 °C

1.524 °C

To achieve the thermocouple accuracies listed above, the USB-2416-4AO should be warmed up for 45 minutes
after the initial power on. The accuracies listed above are only guaranteed if the USB-2416-4AO is housed in its
plastic enclosure.