beautypg.com

Measurement Computing USB-2416-4AO User Manual

Page 31

background image

USB-2416-4AO User's Guide

Specifications

31

Table 11. T type thermocouple accuracy specifications, including CJC measurement error.

All specifications are (±).

A/D data rate Sensor

temperature range

Accuracy error
maximum

Accuracy error
typical

Tempco

(°C/°C)

3750 S/s

-200 °C

2.821 °C

1.676 °C

0.027

0 °C

1.050 °C

0.558 °C

400 °C

0.957 °C

0.595 °C

2000 S/s

-200 °C

2.821 °C

1.676 °C

0.027

0 °C

1.050 °C

0.558 °C

400 °C

0.957 °C

0.595 °C

1000 S/s

-200 °C

2.821 °C

1.676 °C

0.027

0 °C

1.050 °C

0.558 °C

400 °C

0.957 °C

0.595 °C

500 S/s

-200 °C

2.644 °C

1.505 °C

0.027

0 °C

0.979 °C

0.487 °C

400 °C

0.912 °C

0.551 °C

100 S/s

-200 °C

2.644 °C

1.505 °C

0.027

0 °C

0.979 °C

0.487 °C

400 °C

0.912 °C

0.551 °C

60 S/s

-200 °C

2.644 °C

1.505 °C

0.027

0 °C

0.979 °C

0.487 °C

400 °C

0.912 °C

0.551 °C

50 S/s

-200 °C

2.644 °C

1.505 °C

0.027

0 °C

0.979 °C

0.487 °C

400 °C

0.912 °C

0.551 °C

25 S/s

-200 °C

2.644 °C

1.505 °C

0.027

0 °C

0.979 °C

0.487 °C

400 °C

0.912 °C

0.551 °C

10 S/s

-200 °C

2.644 °C

1.505 °C

0.027

0 °C

0.979 °C

0.487 °C

400 °C

0.912 °C

0.551 °C

5 S/s

-200 °C

2.644 °C

1.505 °C

0.027

0 °C

0.979 °C

0.487 °C

400 °C

0.912 °C

0.551 °C

2.5 S/s

-200 °C

2.644 °C

1.505 °C

0.027

0 °C

0.979 °C

0.487 °C

400 °C

0.912 °C

0.551 °C

To achieve the thermocouple accuracies listed above, the USB-2416-4AO should be warmed up for 45 minutes
after the initial power on. The accuracies listed above are only guaranteed if the USB-2416-4AO is housed in its
plastic enclosure.