beautypg.com

Measurement Computing USB-2416-4AO User Manual

Page 30

background image

USB-2416-4AO User's Guide

Specifications

30

Table 10. E type thermocouple accuracy specifications, including CJC measurement error.

All specifications are (±).

A/D data
rate

Sensor temperature
range

Accuracy error
maximum

Accuracy error typical

Tempco
(°C/°C)

3750 S/s

-200 °C

2.413 °C

1.352 °C

0.017

0 °C

1.069 °C

0.551 °C

1000 °C

1.575 °C

1.211 °C

2000 S/s

-200 °C

2.413 °C

1.352 °C

0.017

0 °C

1.069 °C

0.551 °C

1000 °C

1.575 °C

1.211 °C

1000 S/s

-200 °C

2.413 °C

1.352 °C

0.017

0 °C

1.069 °C

0.551 °C

1000 °C

1.575 °C

1.211 °C

500 S/s

-200 °C

2.306 °C

1.244 °C

0.017

0 °C

1.017 °C

0.499 °C

1000 °C

1.539 °C

1.175 °C

100 S/s

-200 °C

2.306 °C

1.244 °C

0.017

0 °C

1.017 °C

0.499 °C

1000 °C

1.539 °C

1.175 °C

60 S/s

-200 °C

2.306 °C

1.244 °C

0.017

0 °C

1.017 °C

0.499 °C

1000 °C

1.539 °C

1.175 °C

50 S/s

-200 °C

2.306 °C

1.244 °C

0.017

0 °C

1.017 °C

0.499 °C

1000 °C

1.539 °C

1.175 °C

25 S/s

-200 °C

2.306 °C

1.244 °C

0.017

0 °C

1.017 °C

0.499 °C

1000 °C

1.539 °C

1.175 °C

10 S/s

-200 °C

2.306 °C

1.244 °C

0.017

0 °C

1.017 °C

0.499 °C

1000 °C

1.539 °C

1.175 °C

5 S/s

-200 °C

2.306 °C

1.244 °C

0.017

0 °C

1.017 °C

0.499 °C

1000 °C

1.539 °C

1.175 °C

2.5 S/s

-200 °C

2.306 °C

1.244 °C

0.017

0 °C

1.017 °C

0.499 °C

1000 °C

1.539 °C

1.175 °C

To achieve the thermocouple accuracies listed above, the USB-2416-4AO should be warmed up for 45 minutes
after the initial power on. The accuracies listed above are only guaranteed if the USB-2416-4AO is housed in its
plastic enclosure.