beautypg.com

Measurement Computing USB-2416 User Manual

Page 25

background image

USB-2416 User's Guide

Specifications

25

Table 5. K type thermocouple accuracy specifications, including CJC measurement error.

All specifications are (±).

A/D data rate Sensor

temperature range

Accuracy error
maximum

Accuracy error
typical

Tempco

(°C/°C)

3750 S/s

-210 °C

2.917 °C

1.699 °C

0.029

0 °C

1.017 °C

0.526 °C

1372 °C

2.478 °C

2.022 °C

2000 S/s

-210 °C

2.917 °C

1.699 °C

0.029

0 °C

1.017 °C

0.526 °C

1372 °C

2.478 °C

2.022 °C

1000 S/s

-210 °C

2.917 °C

1.699 °C

0.029

0 °C

1.017 °C

0.526 °C

1372 °C

2.478 °C

2.022 °C

500 S/s

-210 °C

2.735 °C

1.524 °C

0.029

0 °C

0.948 °C

0.457 °C

1372 °C

2.396 °C

1.941 °C

100 S/s

-210 °C

2.735 °C

1.524 °C

0.029

0 °C

0.948 °C

0.457 °C

1372 °C

2.396 °C

1.941 °C

60 S/s

-210 °C

2.735 °C

1.524 °C

0.029

0 °C

0.948 °C

0.457 °C

1372 °C

2.396 °C

1.941 °C

50 S/s

-210 °C

2.735 °C

1.524 °C

0.029

0 °C

0.948 °C

0.457 °C

1372 °C

2.396 °C

1.941 °C

25 S/s

-210 °C

2.735 °C

1.524 °C

0.029

0 °C

0.948 °C

0.457 °C

1372 °C

2.396 °C

1.941 °C

10 S/s

-210 °C

2.735 °C

1.524 °C

0.029

0 °C

0.948 °C

0.457 °C

1372 °C

2.396 °C

1.941 °C

5 S/s

-210 °C

2.735 °C

1.524 °C

0.029

0 °C

0.948 °C

0.457 °C

1372 °C

2.396 °C

1.941 °C

2.5 S/s

-210 °C

2.735 °C

1.524 °C

0.029

0 °C

0.948 °C

0.457 °C

1372 °C

2.396 °C

1.941 °C

To achieve the thermocouple accuracies listed above, the USB-2416 should be warmed up for 45 minutes after
the initial power on. The accuracies listed above are only guaranteed if the USB-2416 is housed in its plastic
enclosure.