beautypg.com

Screw terminal pinout – differential, Differential connection guidelines – Measurement Computing USB-2416 User Manual

Page 10

background image

USB-2416 User's Guide

Installing the USB-2416

10

Screw terminal pinout – differential

Figure 4. 16-channel differential mode pinout

Differential connection guidelines

 When connecting differential voltage inputs to a "floating" voltage source, make sure there is a DC return

path from each voltage input to ground. You make this path by connecting a resistor from each input to a
GND pin (pins 18, 36, 39, 42, 44, 46, 66, 81, 84). A value of approximately 100 k

Ω can be used for most

applications.

Leave unused input channels either floating or tied to GND (pins 18, 36, 39, 42, 44, 46, 66, 81, 84).

Source impedances should be kept as small as possible to avoid settling time and accuracy errors.

 When configuring thermocouple sensors, keep any stray capacitance as small as possible relative to GND

(pins 18, 36, 39, 42, 44, 46, 66, 81, 84) to avoid settling time and accuracy errors. For thermocouple
channels, do not provide a return path to ground. This is done internally.