beautypg.com

Accuracy, Noise performance, Analog input calibration – Measurement Computing USB-2020 User Manual

Page 19

background image

USB-2020 User's Guide

Specifications

Accuracy

Table 2. DC Accuracy components and specifications. All values are (±)

Range

Gain error
(% of reading)

Offset error
(mV)

INL error
(% of range)

Absolute
accuracy at
Full Scale
(mV)

Gain
temperature
coefficient
(% reading/°C)

Offset
temperature
coefficient
(µV/°C)

±10 V

0.11

5.2

0.0976

35.72

0.0035

30

±5 V

0.11

5.2

0.0488

20.46

0.0035

110

±2 V

0.11

1.1

0.0244

8.18

0.0035

10

±1 V

0.11

1.1

0.0122

4.64

0.0035

25

Noise performance

For the peak-to-peak noise distribution test, a single-ended input channel is connected to AGND at the input

BNC connector and 20,000 data samples are acquired at the maximum rate.

Table 3. Noise performance specifications

Range

Counts

LSBrms

±10 V

5

0.76

±5 V

5

0.76

±2 V

7

1.06

±1 V

7

1.06

Analog input calibration

Table 4. Analog input calibration specifications

Parameter

Specification

Recommended warm-up time 15 minutes min.
Calibration method

Self calibration, with calibration factors for each range stored onboard in non-volatile

memory

Calibration interval

1 year (factory calibration)

19