Analog output

Table 5. Analog output specifications

Parameter

Condition

Specification

Number of channels

2

Resolution

16 bits

Output ranges

Calibrated

±10 V

Output transient

Powered on

Duration: 5 ms
Amplitude: 2 V p-p

Powered off

Duration: 400 ms
Amplitude: 10 V p-p

Differential
non-linearity

16-bit monotonic

±0.35 LSB typ
±1 LSB max

Output current

AOUTx pins

±3.5 mA max (Note 2)

Output coupling

DC

Power on and
reset state

DACs cleared to uncalibrated zero-scale: 0 V, ±50 mV unless the alarm
function is enabled for the output (Note 3)

Alarm functionality

Either or both outputs may be configured to go to defined values when
an Ethernet connection with a host is established or lost.

Output update rate

1000 to 5000 S/s typ, on local network (Note 4)

Slew rate

5 V/µs

Throughput

Software paced

1000 to 5000 S/s typ, on local network (Note 4)

Note 2:

Leave unused AOUTx output channels disconnected.

Note 3:

AOUTx defaults to 0 V whenever the device is powered on or a reset command is issued to the device,
unless the alarm functionality is enabled for the output.

Note 4:

This is the typical throughput when the device and host are both connected by Ethernet to the same
local network. The throughput can vary significantly, and typical throughput is not guaranteed if a
wireless connection is involved or data is sent over the internet.

Table 6. Calibrated absolute accuracy specifications

Range

Absolute accuracy (±LSB)

±10 V

18.7

Table 7. Calibrated absolute accuracy components specifications

Range

% of reading

Offset
(±mV)

Offset tempco
(µV/°C)

Gain tempco
(ppm of range/°C)

±10 V

± 0.024

2.2

30.1

13.2

Table 8. Relative accuracy specifications (±LSB)

Range

Relative accuracy (INL)

±10 V

4.0 typ