beautypg.com

Accuracy, Analog input dc voltage measurement accuracy, Noise performance – Measurement Computing E-1608-OEM User Manual

Page 20: Settling time

background image

20

Accuracy

Analog input DC voltage measurement accuracy

Table 2. DC Accuracy components and specifications. All values are (±)

Range

Gain error
(% of reading)

Offset error
(µV)

INL error
(% of range)

Absolute
accuracy at
Full Scale
(µV)

Gain
temperature
coefficient
(% reading/°C)

Offset
temperature
coefficient
(µV/°C)

±10 V

0.024

915

0.0076

4075

0.0014

47

±5 V

0.024

686

0.0076

2266

0.0014

24

±2 V

0.024

336

0.0076

968

0.0014

10

±1 V

0.024

245

0.0076

561

0.0014

5

Noise performance

For the peak-to-peak noise distribution test, a differential input channel is connected to AGND at the input
terminal block, and 16384 samples are acquired at the maximum rate available at each setting.

Table 3. Noise performance specifications

Range

Counts

LSBrms

±10 V

6

0.91

±5 V

6

0.91

±2 V

7

1.06

±1 V

9

1.36

Settling time

Settling time is defined as the accuracy that can be expected after one conversion when switching from a
channel with a DC input at one extreme of full scale to another channel with a DC input at the other extreme of
full scale. Both input channels are configured for the same input range.

Table 4. Input settling time specifications in µS, typical

Range 4 µS settling accuracy (% FSR) 6 µS settling accuracy (% FSR) 10 µS settling accuracy (% FSR)

±10 V

0.0061

0.0031

0.0015

±5 V

0.0061

0.0031

0.0015

±2 V

0.0061

0.0031

0.0015

±1 V

0.0061

0.0031

0.0015