Product specifications – Ivoclar Vivadent CS User Manual
Page 30

30
9. Product Specifications
9.1 Delivery form
–	 Programat	CS	
–	 Power	cord
–	 Vacuum	hose
–	 Cooling	Tray	“S”
–	 Programat	Firing	tongs
–	 Programat	Firing	Tray	Kit
–	 Automatic	Temperature	Calibration	Set	ATK2	(Test	Set)
–	 USB	data	cable
–	 PrograBase	2	Software	
–	 Service	passport	equipment	
–	 Operating	Instructions
Recommended accessories 
–	 Programat	Accessories	Set
–	 Automatic	Temperature	Checking	Set	2	(ATK	2)
–	 Vacuum	Pump	VP3	easy
9.2 Technical data
Power supply
100 V / 50–60 Hz
110–120 V / 50–60 Hz
200–240 V / 50–60 Hz
Overvoltage	category	II
Contamination	level	2
Tolerated voltage fluctuations
+/- 10%
Max. power consumption
11 A at 100 V
12 A at 110–120 V
8.5 A at 200–240 V
Acceptable	data	for	vacuum	pump	of	other	manufacturers	
Max.	output:		
250 W / max. leakage current 0.75 mA
Final vacuum:
< 50 mbar
Use only tested pumps
Electrical fuses:
100 V / 110–120 V:
250 V / T 15 A (heating circuit)
250 V / T 5 A (vacuum pump)
	
	
200–240 V:
250 V / T 8 A ( heating circuit)
250 V / T 3.15 A ( vacuum pump)
Dimensions of electrical fuses
110–120 V:
Diameter 6.3 x 32 mm
200–240 V:
Diameter 5 x 20 mm
Dimensions	of	the	closed	furnace	
	
Depth: 430 mm / width: 305 mm / 410 mm (with Cooling tray)
Height: 320 mm
Usable size of the firing chamber
Diameter 80 mm
Height 48 mm
Max. firing temperature
1200 °C (2192 °F)
Weight
Furnace base: 12.0 kg
Furnace head: 4.5 kg
Safety information 
The	CS	complies	with	the	following	guidelines:
–	 IEC	1010-1/EN	61010,	Part	1
–	UL	and	cUL	standards
Radio protection / electromagnetic compatibility
EMC tested
9.3 Acceptable operating conditions
Acceptable ambient temperature range:
+5	°C	to	+40	°C	(+41	°F	to	+104	°F)
Acceptable humidity range:
80	%	maximum	relative	humidity	for	temperatures	up	to	31	°C		
(87.8	°	F)	gradually	decreasing	to	50	%	relative	humidity	at	40	°C	
(104	°F);	condensation	excluded.	
Acceptable ambient pressure: 
The	furnace	is	tested	for	use	at	altitudes	of	up	to	2000	m	above	sea	
level.	
9.4 Acceptable transportation and storage conditions
Acceptable temperature range
-20 to +65 °C (-4 °F to +149 °F)
Acceptable humidity range
Max. 80 % relative humidity
Acceptable ambient pressure
500 mbar to 1060 mbar
Use	only	original	packaging	of	the	Programat	CS	together	with	the	
respective	foam	material	for	shipping	purposes.