7.3 Test programs
Press	the	the	key	‘Settings’	and	select	the	desired	test	program	by	
using	the	Cursor-keys.
Vacuum pump test program:
With	this	program,	the	vacuum	performance	of	the	furnace	vacuum	
system	can	be	automatically	tested.	For	that	purpose,	the	achieved	
(minimum)	pressure	in	mbar	is	measured	and	indicated.	If	the	
	pressure	value	is	below	80	mbar,	the	vacuum	performance	of	the	
system	is	adequate.	
Heating muffle test program 
With	this	heater	test,	the	quality	of	the	heating	muffle	can	be	auto-
matically	tested	(duration	approximately	7	minutes).	The	heating	
muffle	test	should	only	be	conducted	with	the	firing	chamber	empty,	
since	any	additional	item	in	the	firing	chamber	(e.g.	firing	tray)		
influences	the	result.	Please	run	the	heating	muffle	test	program	
immediately	after	switching	on	the	furnace	and	before	starting	the	
actual	firing	procedures.	If	the	furnace	is	too	hot,	an	incorrect		
heating	muffle	quality	will	be	indicated.	If	the	quality	of	the	heating	
elements	drops	below	50%,	changing	the	heating	element	is		
strongly	recommended.
Keypad test program 
With	every	keystroke	a	short	beep	sounds.	The	keypad	test	is	ended	
by	pressing	the	ESC	key.
Cleaning program 
The	heating	muffle	is	‚cleaned‘	using	the	cleaning	program	(duration	
approximately	17	minutes). 
7.4 Stand-by
We	recommend	keeping	the	furnace	head	closed,	especially	if	the	
temperature	drops	below	150	°C	(302	°F).
7.5 Dehumidification program
The	condensation	of	water	in	the	insulation	of	the	firing	chamber	
and	the	vacuum	pump	will	result	in	a	lower	vacuum	and	thus		
impaired	firing	results.	For	that	reason,	the	furnace	head	should	be	
kept	closed	when	the	furnace	is	switched	off,	in	order	to	prevent	the	
absorption	of	humidity.	Start	the	dehumidification	program	if		
required	(humidity	in	the	insulation).	
7.6 Temperature calibration
1. Select the calibration program.
2. Remove the firing plate from
the	furnace	using	the	furnace	
tongues	and	place	it	on	the	
cooling	tray.	
3. Carefully grip the upper part
of	the	ATK	2	using	the	furnace	
tongs	(Caution:	Fracture	risk	of	
the	ceramic)	and	insert	it	into	
the	holes	designated	for	this		
purpose	until	it	snaps	into		
place.	The	orientation	of	the	
calibration	sample	(left	or	right)	
is	not	important.	
4. If necessary, use the furnace
tongs	to	apply	slight	pressure	
to	the	center	of		the	calibration	
base	until	the	calibration		
sample	clicks	into	place.		
Observe	the	corresponding	
markings.	
5. Start the calibration program
6. At the end of the program,
open	the	furnace	head	and	
carefully	remove	the	ATK	2	
using	the	furnace	tongs	and	
place	it	on	the	cooling	tray	to	
allow	it	to	cool.
7. Replace the firing plate using the furnace tongs.
8. Close the furnace head and select a firing program.
9. The ATK 2 can only be used once. Use a new calibration set for
the next calibration procedure.