Maintenance, cleaning, and diagnosis – Ivoclar Vivadent CS User Manual
Page 25

25
7. Maintenance, Cleaning, and Diagnosis
This	chapter	describes	the	user	maintenance	and	cleaning	procedures	
for	the	Programat	CS.	All	the	other	tasks	must	be	performed	by		
qualified	service	personnel	at	a	certified	Ivoclar	Vivadent	Service		
Center.			
7.1 Monitoring and maintenance
The	time	for	these	maintenance	procedures	depends	on	the		
frequency	of	use	and	the	working	habits	of	the	users.	For	that		
reason,	the	recommended	times	are	only	approximations.
This furnace has been developed for typical use in dental
laboratories.	If	the	product	is	used	in	a	production		
enterprise,	for	industrial	applications,	and	for	continuous	
use,	premature	ageing	of	the	expendable	parts	has	to	be	
expected.
The	expendable	parts	are	as	follows:
–	 Heating	muffle
–		Insulation	material
Expendable	parts	are	not	covered	by	the	warranty.	Please	also		
observe	the	shorter	service	and	maintenance	intervals.
If the furnace head is replaced, the furnace must be calibrated.
7.2 Cleaning
The furnace may only be cleaned when it is cool, since
there is a burn hazard. Do not use any cleaning solutions.
The following parts have to be cleaned from time to time:
*Never clean with compressed air!
What
Part
When
Check all plug-in connections for correct fit
Var. external connections
weekly
Check if the furnace head opens smoothly and without excessive noise.
Opening mechanism
monthly
Check if the thermocouple is straight and in the right place.
Thermocouple (4)
weekly
Check	the	insulation	for	cracks	and	damages.	If	the	insulation	is	worn	down	
it	has	to	be	replaced	by	a	certified	Ivoclar	Vivadent	Service	Center.	Fine	hair-
line	cracks	on	the	surface	of	the	insulation	are	harmless	and	do	not		
influence	the	function	of	the	furnace	in	a	negative	fashion.	
Insulation (3)
monthly
Check	if	the	sealing	rims	of	the	furnace	head	and	the	furnace	base	are	clean	
and	undamaged.	
Sealing	rims	of	the	furnace	head	(2)	
and	the	furnace	base	(1)
weekly
Check	the	keypad	for	visible	damage.	If	the	keypad	is	damaged,	it	has	to	be	
replaced	by	a	certified	Ivoclar	Vivadent	Service	Center.
Keypad (10)
weekly
Check	temperature.	
Use	the	temperature	checking	set	to	check	and	adjust	the	temperature	in	
the	furnace.
Firing chamber
twice a year
Check	the	quartz	glass	cylinder	to	make	sure	the	quartz	glass	is	not	
	defective.
Firing chamber
daily
Item:
Frequency:
Cleaning material:
Housing	(9)	and	furnace	
head	(25)	
if required
soft, dry cloth
Keypad (10)
weekly
soft, dry cloth
Cooling tray (34)
daily
cleaning brush *
Insulation (3)
daily
cleaning brush *
Sealing	rim	of	the	
	furnace	head	(2)	and	
sealing	surface	(1)
daily
cleaning	brush	and	a	
soft	cloth	*
