beautypg.com

Optional – Ivoclar Vivadent PS Empress CAD Chairside User Manual

Page 29

background image

29

Fabrication of IPS Empress

®

CAD Restorations

– V

eneers / Cr

owns – stained / glazed

Stained and glazed IPS Empress CAD crown and veneer

Optional

Subsequent adjustments

After completion, additional adjustments (e.g. contact points) may become necessary. For that purpose,
IPS Empress Add-On 770 °C / 1418 °F is available.

Procedure:
– Before the adjustment, clean the restoration thoroughly from dirt and grease using a steam jet.
– Mix IPS Empress Add-On 770 °C / 1418 °F with IPS Empress Esthetic Veneer Build-Up Liquid to a workable consistency.

Make sure that the add-on material and liquid are evenly mixed so that an optimum firing result can be achieved.

– Apply the mixed add-on material on the cleaned, desired areas using a brush and slightly blot them with an absorbent

cloth.

– Position the adjusted restoration on the firing tray and fire it in the ceramic furnace.
– After firing, manually polish the supplemented areas to a high gloss.

Firing parameters for the corrective firing

IPS Empress CAD with
IPS Empress Add-On 770 °C / 1418 °F

B

°C / °F

S

min.

t

°C/min. / °F/min.

T

°C / °F

H

min.

V

1

°C / °F

V

2

°C / °F

Corrective firing

403

04:00

60

770

02:00

450

769

B = Stand-by temperature °C / °F

S = Closing time/min.

t = Temperature increase rate °C/min. / °F/min.

T = Firing temperature °C / °F

H= Holding time/min.

V

1

= Vacuum-on temperature °C / °F

V

2

= Vacuum-off temperature °C / °F

This manual is related to the following products: