beautypg.com

5 calibration methods, 6 noise calibration with scream – Guralp Systems CMG-6T User Manual

Page 23

background image

Operator's Guide

3.5 Calibration methods

Velocity sensors such as the 6T are not sensitive to constant DC levels,
either as a result of their design or because of an interposed high-pass

filter. Instead, three common calibration techniques are used.

Injecting a step current allows the system response to be
determined in the time domain. The amplitude and phase

response can then be calculated using a Fourier transform.
Because the input signal has predominantly low-frequency

components, this method generally gives poor results. However,
it is simple enough to be performed daily.

Injecting a sinusoidal current of known amplitude and

frequency allows the system response to be determined at a spot
frequency. However, before the calibration measurement can be

made the system must be allowed to reach a steady state; for low
frequencies, this may take a long time. In addition, several

measurements must be made to determine the response over the
full frequency spectrum.

Injecting white noise into the calibration coil gives the response

of the whole system, which can be measured using a spectrum
analyser.

You can calibrate a 6T sensor using any of these methods, using the

signal generator built in to all Güralp Systems' digitisers.

3.6 Noise calibration with Scream!

The most convenient way to calibrate a 6TD instrument is to use a

CMG-DM24's pseudo-random broadband noise generator along with
Scream!'s noise calibration extension. The extension is part of the

standard distribution of Scream!, and contains all the algorithms
needed to determine the complete sensor response in a single

experiment.

Information on other calibration methods is available on the Güralp
Systems Web site.

1. In Scream!'s main window, right-click on the digitizer's icon and

select Control.... Open the Calibration pane.

2. Select the calibration channel corresponding to the instrument,

and choose Broadband Noise. Select the component you wish

to calibrate, together with a suitable duration and amplitude,
and click Inject now. A new data stream ending Cn (n = 0 – 7)

July 2011

23