Groth 3000 Series User Manual
Page 16

13
RECOMMENDED TESTING AND SETTING PROCEDURE
 
1.0) Equipment 
 
1.1) A high pressure air (or inert gas) supply 
system capable of maintaining supply pressure 
at 0-200 PSI. Flow may be controlled by a 
reducing regulator or valve. 
1.2) A low pressure air (gas) supply system
capable of maintaining sense pressure at 1/2" 
WC to 10 PSI. Pressure may be controlled by 
a regulator or needle valve and a small 
accumulator. If a needle valve is used, it will 
require frequent adjustment because the 
regulator consumes a significant volume of air 
when operating. 
1.3) A means of indicating or measuring air flow
with a capacity of approximately 30 SCFH.
1.4) A means of soap bubble shell testing the
assembly.
 
2.0) Test procedure and Acceptance Criteria. 
 
2.1) Connect the high pressure air supply to the 
regulator inlet port.
2.2) Connect the low pressure air supply to the
regulator sense port.
2.3) Connect a reduced size manual shut-off valve
to the outlet port.
2.4) Connect a small diameter hose or plastic tube
to the inlet of the flow indicator or flow meter. 
Use a hand held port cover to attach the flow 
meter to the regulator outlet port. Never 
connect the flow meter rigidly to the regulator 
unless the flow meter is rated for full air 
supply pressure. 
2.5) Connect a small diameter hose or plastic tube
to the outlet of the flow indicating device so 
the end may be immersed in a container of 
water. 
2.6) Close the outlet valve and set the supply air
pressure to the maximum available, but not 
greater than 300 PSI. Soap bubble test the 
entire assembly at this pressure. Open the 
outlet valve and reduce the supply until a 
minimum audible flow is observed. 
 
 
2.7) Adjust the sense pressure to the specified 
regulator set pressure. Adjust the regulator 
screw to shut off (no audible flow). Now 
increase the sense pressure to 2 times the 
specified set pressure or 1.5 times the 
specified tank MAWP, whichever is greater. 
Soap bubble test the actuator housing. 
2.8) Increase the supply pressure to the specified
system pressure (100 PSI if not specified).
2.9) Slowly decrease the sense pressure and adjust
the regulator screw until the air flow is 5-20 
SCFH at the specified setting. Lock the 
adjusting screw at this setting. 
2.10) While immersing the outlet tube in water,
increase the sense pressure until the regulator 
is bubble tight. If this pressure exceeds 15% 
or 2" WC greater than the set pressure, 
examine the seat and O-Ring. Do not adjust 
the regulator screw to improve bubble 
tightness. 
2.11) Remove the bonnet vent. Shut off the supply
pressure. Slowly decrease the sense pressure 
to zero, while observing the stroke of the 
piston thru the bonnet vent port. The stroke 
must be 0.18" minimum. Replace the bonnet 
vent. 
 
3.0) Test report data. 
 
3.1) The following data should be recorded for 
future reference:
Model, Tag & Serial numbers
Test
media
Body & Actuator shell test pressure
Supply
pressure
Set
pressure
Bubble tight pressure
 
 
 
 
 
 
 
 
 
 
 
