Cmd_query_osci_trigger_filter_results, Cmd_query_osci_main_filter_results – GBS Elektronik MCA-527 Firmware Commands User Manual
Page 52
MCA527
Name
CMD_QUERY_OSCI_TRIGGER_FILTER_RESULTS
Compatibility
New MCA527 command.
Execution right
Necessary
Description
The command reads data that are calculated from the oscilloscope data during the last call of the
command CMD_QUERY_OSCI_SCREEN_EX if the corresponding flag is set.
Range: 0:
results [0 … 359]
1:
results [360 … 719]
Format
integer
integer
integer
integer
integer
integer
Parameter
Preamble
command
range
0
0
end flag
Byte string (HEX)
A5
5A
2A
01
range
l
range
h
00
00
00
00
B9
9B
Result data array
Byte offset 0
Applied trigger threshold [ * 0.00006103515625]
long
Byte offset 4
Results [ * 0.00006103515625]
long [360]
Byte offset 1444
Command flag and parameters
8 bytes
Byte offset 1452
Checksum
unsigned short
Remarks
In order to read all data, the command has to be called twice. The first time with range=0 and the
second time with range=1.
The oscilloscope mode is described in document: “MCA527 Oscilloscope Mode”.
Name
CMD_QUERY_OSCI_MAIN_FILTER_RESULTS
Compatibility
New MCA527 command.
Execution right
Necessary
Description
The command reads data that are calculated from the oscilloscope data during the last call of the
command CMD_QUERY_OSCI_SCREEN_EX if the corresponding flag is set.
Format
integer
integer
integer
integer
integer
integer
Parameter
Preamble
command
0
0
0
end flag
Byte string (HEX)
A5
5A
2B
01
00
00
00
00
00
00
B9
9B
Result data array
Byte offset 0
Results
short [720]
Byte offset 1440
Jitter
short
Byte offset 1442
Command flag and parameters
8 bytes
Byte offset 1450
Checksum
unsigned short
Remarks
The oscilloscope mode is described in document: “MCA527 Oscilloscope Mode”.
55 The value is up to now unused. It is always 0.
52