Cmd_set_bias, Cmd_set_pin5_current_source – GBS Elektronik MCA-527 Firmware Commands User Manual
Page 18
MCA527
Command name
CMD_SET_BIAS
Compatibility
Compatible to the MCA166 command.
Execution right
Necessary
Description
The command sets the detector high voltage and controls the HV inhibit input.
hv:
0 ... 3600 V
i = 0
Inhibit off
i = 1
”Canberra HPGe mode”,
HV shut down if inhibit input < 0.5V
i = 2
“DSG HPGe mode”,
HV shut down if inhibit input < 0.5V
i = -1
”Ortec HPGe mode”,
HV shut down if inhibit input ≥ 5V
Format
integer
integer
integer
integer
integer
integer
Parameter
preamble
command
hv
i
0
end flag
Byte string (HEX)
A5
5A
4F
00
hv
l
hv
h
i
l
i
h
00
00
B9
9B
Remarks
The Lite version provides no HV inhibit input. Setting the HV inhibit input on is ignored.
The OEM version provides neither HV nor HV inhibit input. All settings are ignored.
If the connected detector supplies detector information, the MCA527 compares the polarity of the
detector with that of the power module. If the polarities are different, all settings are ignored.
Command name
CMD_SET_PIN5_CURRENT_SOURCE
Compatibility
New MCA527 command
Execution right
Necessary
Description
The command switches the current source (cs) on SUB-D9 pin5 on (= 1) or off (= 0).
Format
integer
integer
integer
long
integer
Parameter
preamble
command
cs
0
end flag
Byte string (HEX)
A5
5A
19
01
cs
00
00
00
00
00
B9
9B
Remarks
The SUB-D9 pin5 is supported by the full version and partially by OEM version.
20 Mode 1 and 2 are identical. The difference is made by reasons of legacy.
18