Description of the MCA527 Firmware Commands
Command name
CMD_SET_EXTENSION_PULSER_WIDTH
Compatibility
New MCA527 command.
Execution right
Necessary
Description
The command sets the pulser width (w) for the parts (part) of the extension port.
part:
1 = part B (pulser2)
3 = part D (pulser1)
w:
The pulser width must be smaller than the pulser period (see previous command). If the
parameter does not meet this condition, the command will return “invalid parameter”.
Pulser 1:
1 … 4294967294 * 10 ns
Pulser 2
1 … 4294966
* 10 µs
Format
integer
integer
integer
long
integer
Parameter
preamble
command
part
w
end flag
Byte string (HEX)
A5
5A
1D
01
part
00
w
l
...
...
w
h
B9
9B
Command name
CMD_SET_EXTENSION_RS232
Compatibility
New MCA527 command.
Execution right
Necessary
Description
The command configures the RS232 interface of the extension port.
div:
0x0001 … 0xFFFF (baud rate = 6250000 / div, see predefined constants in mca_comm.h)
flags:
bits 1, 0: 00
5-bit word
01
6-bit word
10
7-bit word
11
8-bit word
bit 2:
0
1 stop bit
1
2 stop bits for non-5-bit word length or 1½ stop bits for 5-bit word
length
bit 3:
0
parity not transmitted or checked
1
transmit and check parity
bit 4:
0
Odd parity
1
Even parity
Format
integer
integer
integer
integer
integer
integer
Parameter
preamble
command
div
flags
0
end flag
Byte string (HEX)
A5
5A
1E
01
div
l
div
h
flags
00
00
00
B9
9B
Command name
CMD_CLEAR_EXTENSION_RS232_TX
Compatibility
New MCA527 command.
Execution right
Necessary
Description
The command clears all bytes which has been written to the RS232 transfer buffer.
Format
integer
integer
integer
long
integer
Parameter
preamble
command
0
0
end flag
Byte string (HEX)
A5
5A
1F
01
00
00
00
00
00
00
B9
9B
25