Description of the mca527 firmware commands – GBS Elektronik MCA-527 Firmware Commands User Manual
Page 29
Description of the MCA527 Firmware Commands
Command name
CMD_QUERY_POWER (Continuation)
Result data array
Byte offset 72
+12V primary current [mA] at stop
unsigned long
Byte offset 76
-12V primary current [mA] at stop
unsigned long
Byte offset 80
+24V primary current [mA] at stop
unsigned long
Byte offset 84
-24V primary current [mA] at stop
unsigned long
Byte offset 88
Battery voltage [mV] at stop
unsigned long At the MCA527Micro this data
corresponds to the USB input
voltage.
Byte offset 92
HV [ * 1.2 V] at stop
unsigned long
Byte offset 96
ADC correction offset on SUB-D9 pin3 [LSB]
char
Byte offset 97
Gain correction factor on SUB-D9 pin3
char
factor = 0.001 * value + 1
Byte offset 98
unused
2 bytes
Byte offset 100
+12V actual value [ * 0.0625 V] at stop
unsigned char
Byte offset 101
-12V actual value [ * 0.0625 V] at stop
unsigned char
Byte offset 102
+24V actual value [ * 0.125 V] at stop
unsigned char
Byte offset 103
-24V actual value [ * 0.125 V] at stop
unsigned char
Byte offset 104
Voltage on SUB-D9 pin3 [ * 0.3125 mV] at stop
unsigned short
Byte offset 106
Command flag and parameters
8 bytes
Byte offset 114
Voltage on SUB-D9 pin5 [ * 0.3125 mV] at stop
unsigned short
Byte offset 116
Charger current [mA] at stop
unsigned long
Byte offset 120
unused
6 bytes
Byte offset 126
Checksum
unsigned short
Byte offset 128
MCA state
integer value
STATE_READY
= 1
STATE_RUN
= 2
STATE_SUSPEND
= 3
STATE_FINISH
= 4
STATE_STOP
= 5
STATE_FAIL
= 6
STATE_WAIT_FOR_TRIGGER
= 7
Byte offset 130
unused
2 bytes
29