Cmd_set_pur, Cmd_set_fast, Cmd_set_slow – GBS Elektronik MCA-527 Firmware Commands User Manual
Page 14

MCA527
Command name
CMD_SET_PUR
Compatibility
Identical to the MCA166 command.
Execution right
Necessary
Description
The command sets the pile up rejection.
pur = 0 turns PUR off
pur
≠
0 turns PUR on
The command is ignored and responds with an error value if a measurement is still running.
Format
integer
integer
integer
long
integer
Parameter
preamble
command
pur
0
end flag
Byte string (HEX)
A5
5A
53
00
pur
00
00
00
00
00
B9
9B
Command name
CMD_SET_FAST
Compatibility
Compatible to the MCA166 command. The command is meaningless for the MCA527. It has only
been retained for compatibility with older software
Execution right
Necessary
Description
This command sets the fast discriminator threshold (0 ... 2499, default 400).
Format
integer
integer
integer
long
integer
Parameter
Preamble
command
fast
0
end flag
Byte string (HEX)
A5
5A
50
00
fast
l
fast
h
00
00
00
00
B9
9B
Command name
CMD_SET_SLOW
Compatibility
Compatible to the MCA166 command. The command is meaningless for the MCA527. It has only
been retained for compatibility with older software
Execution right
Necessary
Description
This command sets the slow discriminator threshold (0 ... 2499, default 400).
Format
integer
integer
integer
long
integer
Parameter
Preamble
command
slow
0
end flag
Byte string (HEX)
A5
5A
51
00
slow
l
slow
h
00
00
00
00
B9
9B
14