Cmd_query_state527_ex2, Description of the mca527 firmware commands – GBS Elektronik MCA-527 Firmware Commands User Manual
Page 41

Description of the MCA527 Firmware Commands
Command name
CMD_QUERY_STATE527_EX2
Compatibility
New MCA527 command (since firmware version 14.00).
Execution right
Not necessary
Description
The command reads the state.
Format
integer
integer
integer
long
integer
Parameter
preamble
command
0
0
end flag
Byte string (HEX)
A5
5A
2F
01
00
00
00
00
00
00
B9
9B
Result data array
Byte offset 0
AHRC group 0 width
unsigned long
Byte offset 4
AHRC group 1 width
unsigned long
Byte offset 8
AHRC group 2 width
unsigned long
Byte offset 12
AHRC group 3 width
unsigned long
Byte offset 16
AHRC group 4 width
unsigned long
Byte offset 20
AHRC group 5 width
unsigned long
Byte offset 24
AHRC group 6 width
unsigned long
Byte offset 28
AHRC group 7 width
unsigned long
Byte offset 32
AHRC group 8 width
unsigned long
Byte offset 36
AHRC group 9 width
unsigned long
Byte offset 40
AHRC trigger threshold
unsigned short
Byte offset 42
unused
2 bytes
Byte offset 44
Time window 0 width for gating mode 'sort by time'
unsigned long
Byte offset 48
Time window 1 width for gating mode 'sort by time'
unsigned long
Byte offset 52
Time window 2 width for gating mode 'sort by time'
unsigned long
Byte offset 56
Time window 3 width for gating mode 'sort by time'
unsigned long
Byte offset 60
Time window 4 width for gating mode 'sort by time'
unsigned long
Byte offset 64
Time window 5 width for gating mode 'sort by time'
unsigned long
Byte offset 68
Time window 6 width for gating mode 'sort by time'
unsigned long
Byte offset 72
Time window 7 width for gating mode 'sort by time'
unsigned long
Byte offset 76
unused
30 bytes
Byte offset 106
Command flag and parameters
8 bytes
Byte offset 114
unused
12 bytes
Byte offset 126
Checksum
unsigned short
Byte offset 128
MCA state
unsigned short See CMD_QUERY_POWER
Byte offset 130
unused
2 bytes
50 Since firmware version 14.02.
41