Description of the MCA527 Firmware Commands
Command name
CMD_SET_FLAT_TOP_TIME
Compatibility
New MCA527 command.
Execution right
Necessary
Description
The command sets the flattop time (ftt = 0 … 50 * 0.1 µs).
The command is ignored and responds with an error value if a measurement is still running.
Format
integer
integer
integer
long
integer
Parameter
preamble
command
ftt
0
end flag
Byte string (HEX)
A5
5A
13
01
ftt
00
00
00
00
00
B9
9B
Command name
CMD_SET_JITTER_CORRECTION
Compatibility
New MCA527 command.
Execution right
Necessary
Description
The command sets the jitter correction.
jc = 0
turns jitter correction off
jc
≠
0
turns jitter correction on
The command is not standardly handled. It is only handled if the flag
FEATURES_JITTER_CORRECTION is set within the value “MCA features” that is returned by
CMD_QUERY_STATE527, byte offset 8.
The command is ignored and responds with an error value if a measurement is still running.
Format
integer
integer
integer
long
integer
Parameter
preamble
command
jc
0
end flag
Byte string (HEX)
A5
5A
16
01
jc
00
00
00
00
00
B9
9B
Command name
CMD_SET_BASELINE_RESTORING
Compatibility
New MCA527 command.
Execution right
Necessary
Description
The command sets the baseline restorer.
blr = 0 off
blr = 1
1/1
blr = 2
1/2
blr = 3
1/4
blr = 4
1/8
blr = 5
1/16
blr = 6
1/32
The command is not standardly handled. It is only handled if the flag
FEATURES_ADJUSTABLE_BASELINE_RESTORER is set within the value “MCA features” that is
returned by CMD_QUERY_STATE527, byte offset 8.
The command is ignored and responds with an error value if a measurement is still running.
Format
integer
integer
integer
long
integer
Parameter
preamble
command
blr
0
end flag
Byte string (HEX)
A5
5A
18
01
blr
00
00
00
00
00
B9
9B