Cmd_query_uf6_rois – GBS Elektronik MCA-527 Firmware Commands User Manual
Page 48

MCA527
Name
CMD_QUERY_UF6_ROIS
Compatibility
Identical to the MCA166 command.
Execution right
Not necessary
Description
The command reads begin and end of up to three ROIs that can be set by the command
CMD_SET_UF6_ROIS. The name of the command refers to its original usage. See also the
command CMD_QUERY_UF6_INFO.
Format
integer
integer
integer
long
integer
Parameter
Preamble
command
0
0
end flag
Byte string (HEX)
A5
5A
65
00
00
00
00
00
00
00
B9
9B
Result data array
Byte offset 0
1
st
ROI begin
unsigned long
Byte offset 4
1
st
ROI end
unsigned long
Byte offset 8
2
nd
ROI begin
unsigned long
Byte offset 12
2
nd
ROI end
unsigned long
Byte offset 16
3
rd
ROI begin
unsigned long
Byte offset 20
3
rd
ROI end
unsigned long
Byte offset 24
unused
82 bytes
Byte offset 106
Command flag and parameters
8 bytes
Byte offset 114
unused
12 bytes
Byte offset 126
Checksum
unsigned short
Byte offset 128
unused
4 bytes
48