Fluke Biomedical QED 6 User Manual
Page 60

QED 6
Users Guide
2-34
Fluke Biomedical
Control #:
Serial #:
Model #:
Mfr.
Location:
Technician:
Date:
PROGRAM NAME: PROG 5
Settin
g
Actual Limit +/- 5
%
10
J
10.3
J
100
J 10.3
J
200
J 199.0
J
300
J 300.0
J
360
J 360.0
J
Char
g
e Time: 5 sec
19
8
.3 J
Qwave Sync Time 21.0
Rwave Sync Time 50.0
Peak Volta
g
e:
2270 V
Peak Current:
45 A
OVER Volta
g
e:
2 V
OVER Current:
2 A
ECG/Performance Waves
30 BPM
120 BPM
240 BPM
2 Hz S
q
.125 Hz S
q
10 Hz Sin
40 Hz Sin
50 Hz Sin
60 Hz Sin
100 Hz Sin
2 Hz Tri
60 BPM Pls
Vfib
fcf051.eps
Figure 2-11. Automatic Sequence Output with Header
See also other documents in the category Fluke Biomedical Equipment:
- 7000DP Impulse Supplement (12 pages)
- 7000DP Impulse Getting Started (36 pages)
- 7000DP Impulse (92 pages)
- 06-526-2200 (26 pages)
- 07-417 (16 pages)
- 07-443 (14 pages)
- 07-451 (8 pages)
- 07-487 (12 pages)
- 07-453 (24 pages)
- 07-555 (8 pages)
- 07-553 (10 pages)
- 07-605-7777 (8 pages)
- 07-444 (46 pages)
- 07-618 (10 pages)
- 07-600 (6 pages)
- 07-591 (8 pages)
- SigmaPace 1000 (154 pages)
- 07-653 (8 pages)
- 07-633 (20 pages)
- 07-649 (14 pages)
- 07-661-7662 (12 pages)
- 07-645 (14 pages)
- 10100AT (80 pages)
- 07-644 (10 pages)
- 18-203 (4 pages)
- 07-621 (12 pages)
- 07-647 (12 pages)
- 18-207 (8 pages)
- 18-216-1000 (10 pages)
- 18-220 (10 pages)
- 18-228 (8 pages)
- 18-223 (18 pages)
- 18-229-1313 (6 pages)
- 18-250 (8 pages)
- 18-252 (6 pages)
- 18-303 (6 pages)
- 35035 (19 pages)
- 6000-528 (18 pages)
- 35080B (70 pages)
- 6000-529 (16 pages)
- 35080M (64 pages)
- 6000-530B (16 pages)
- 57-436 (12 pages)
- 57-440 (12 pages)
- 57-402 (32 pages)