Operation
Programming an Automatic Test Sequence
2
2-27
To program an automatic test sequence, carry out the following steps and those
listed under individual headings, below:
1. From
Main Menu 2
, press the
AUTO
softkey to display the
Autosequences
menu:
fcf034.eps
2. Press the
PROG
softkey to access the individual programs to be modified.
fcf035.eps
Program Selection
1. Select the automatic test sequence to be modified by pressing the
+
or
-
softkeys to increase / decrease program numbers.
Available programs are named
PROG
and numbered 0-27, but you can
press the
NAME
softkey and modify the program name and number.
2. Press the
SEL
softkey to confirm the program selected for modification
and to attach a header.
You are asked if you wish to attach a header to the data to be output after
the test sequence has been run.
fcf036.eps
3. Press the
SEL
softkey to toggle between
Yes
and
No
.
- 7000DP Impulse Supplement (12 pages)
- 7000DP Impulse Getting Started (36 pages)
- 7000DP Impulse (92 pages)
- 06-526-2200 (26 pages)
- 07-417 (16 pages)
- 07-443 (14 pages)
- 07-451 (8 pages)
- 07-487 (12 pages)
- 07-453 (24 pages)
- 07-555 (8 pages)
- 07-553 (10 pages)
- 07-605-7777 (8 pages)
- 07-444 (46 pages)
- 07-618 (10 pages)
- 07-600 (6 pages)
- 07-591 (8 pages)
- SigmaPace 1000 (154 pages)
- 07-653 (8 pages)
- 07-633 (20 pages)
- 07-649 (14 pages)
- 07-661-7662 (12 pages)
- 07-645 (14 pages)
- 10100AT (80 pages)
- 07-644 (10 pages)
- 18-203 (4 pages)
- 07-621 (12 pages)
- 07-647 (12 pages)
- 18-207 (8 pages)
- 18-216-1000 (10 pages)
- 18-220 (10 pages)
- 18-228 (8 pages)
- 18-223 (18 pages)
- 18-229-1313 (6 pages)
- 18-250 (8 pages)
- 18-252 (6 pages)
- 18-303 (6 pages)
- 35035 (19 pages)
- 6000-528 (18 pages)
- 35080B (70 pages)
- 6000-529 (16 pages)
- 35080M (64 pages)
- 6000-530B (16 pages)
- 57-436 (12 pages)
- 57-440 (12 pages)
- 57-402 (32 pages)